NormaChem 200

  • Fully automated
  • Random access
  • Benchtop, small footprint

Technical Data

Throughput: 200 tests/hour, up to 275 tests/hour with ISE (optional)
Test Mode: End point, Kinetics, Immunoturbidimetry, Fixed time, Monochromatic, Bi-chromatic & Multi-standard Analysis
Linear & Non-linear Calibration
Test Setting: Programmable for single-parameter, multi-parameter, combined parameter, batch test
Test Range: Clinical Biochemistry, Turbidimetry, ISE, Protein, Drug abuse

Sampling System

Sample Tray: 44 sample positions for primary tubes,test tubes and various sample cups
Whole sample tray can be removed and replaced
Any sample position can be selected as emergency test position
Sample Volume: 3µL~ 30µL, programmed by 0.1µL step
Sample Handling: Dilute and retest automatically when sample’s concentration is beyond the limit
Sample probe: Liquid level detection and collision protection
Probe cleaning: Automatic washing inside and outside, 
carry over <0.1%
Built-in barcode reader: Optional

Reagent System

Reagent tray: 44 positions for Single & Dual reagents, reagent positions can be selected freely
Refrigeration: Refrigerated reagent compartment, 24 hours non-stop cooling 
Reagent bottle volume: 35mL and 20mL, sloped bottom design
Reagent volume: 150µL ~300µL, programmed by 0.1µL step
Reagent probe: Automatic liquid level detection and collision protection
Reagent pre-heating function,inventory checking
Real time monitoring of reagent residual volume and auto alarm.
Carry over 0.1%

Reaction System

Reaction tray: 48 reaction cuvettes with 6mm optical path, good UV penetrability
Reaction technology: Back-Dividing-Light
Reaction volume: 250µL ~300µL
Temperature: 37℃, with ± 0.1℃ fluctuation
Mixing system: Independent mixer
Cleaning: Automatic cuvette cleaning and complete drainage to reduce carry over
Liquid detection: Detergent, distilled water, waste liquid automatic detection and alarm

Optical System

Light source: Halogen lamp 6V/10W, life span≥ 3000 hours
Wavelength: 340nm, 405nm, 450nm, 492nm, 505nm, 546nm, 578nm, 620nm, 670nm, 700nm
Optical accuracy: ±2nm
Absorbance range: 0.000Abs~3.000Abs

Calibration & Quality Control

Calibration mode: Linear (one-point, two-point and multi-point), Logit-Log4P, Logit-Log5P, Spline, Exponential ,Polynomial, Parabola
QC rules: Westgard Multi-rule
QC mode: Intra day & Inter day QC

Working Conditions

Power: 110-230V, 50/60Hz
Power consumption: ≤400VA
Environmental temperature: 15℃~30℃
Dimensions: L:740 mm W: 520 mm H:467 mm
Weight: 65 kg